Starter Kit Instruction Manual Maintenance 42
8.3 TCU151 temperature c on-
troller interlock setting
8.3.1 Generalities
The interlock is located on the main board.
Note: By default, the interlock is set to NC (Nor-
mally Close) contact.
8.3.2 Before beggining interlock set-
ting
[1] Switch the instrument OFF.
[2] Unplug the power cord.
[3] Remove the cover (1) to acces the main
board (2).
8.3.3 Needed material
1 set of screwdriver.
1 tweezer.
8.3.4 Procedure interlock setting
To set the type of interlock, proceed as follows:
[1] Locate the J9 Interlock connector.
[2] Place the associated jumper according to
the type of desired interlock.
Note: The common pin is on the center (4)
and by placing the jumper on a side defines
the interlock type as:
for a NC contact: jumper on pins (4) and
(5)
for a NO contact: jumper on pins (3)
and (4)
2
1
T
e
m
p
e
r
a
t
u
r
e
C
o
n
t
r
o
l
l
e
r
A
l
a
r
m
R
e
s
e
t
E
x
t
e
r
n
a
l
I
n
t
e
r
n
a
l
R
e
a
l
I
R
e
a
l
C
Se
t
t
i
n
g
C
S
e
t
t
i
n
g
+
I
S
e
t
t
i
n
g
I
+
TCU151
Fig.31: Opening the TCU151 cover
3
4
5
Fig.32: J9 Interlo ck connector
Vue de la page 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 59 60

Commentaires sur ces manuels

Pas de commentaire